MPT120 120 Watt Medical Power Supplies

Universal AC input

Low leakage current ≤180 umA

Protections: Short circuit / Overload / Over voltage

Fixed switching frequency at 45KHz

Model Number

Output Voltage

Output Amps

Minimum Load Requirement

Load Regulation

Line Regulation

Ripple and Noise

MPT120-A 5vdc/ 12vdc/ -5vdc 10 Amps/ 4.8 Amps/ .6 Amps

20%

±1.0%/ ±3.5%/ ±1.0%

±0.5%/±1.5%/±0.5%

80mV pk-pk/120mV pk-pk/80mV pk-pk (note 1)

MPT120-B 5vdc/ 12vdc/ -12vdc 10 Amps/ 4.4 Amps/ .6 Amps

20%

±1.0%/ ±3.5%/ ±1.0%

±0.5%/±1.5%/±0.5%

80mV pk-pk/120mV pk-pk/80mV pk-pk (note 1)

MPT120-C 5vdc/ 15vdc/ -15vdc 10 Amps/ 4 Amps/ .6 Amps

20%

±1.0%/ ±3.2%/ ±1.0%

±0.5%/±2.0%/±0.5%

80mV pk-pk/150mV pk-pk/80mV pk-pk (note 1)

MPT120-D 5vdc/ 24vdc/ 12vdc 10 Amps/ 2.2 Amps/ .6 Amps

20%

±1.0%/ +4, -3%/ ±1.0%

±0.5%/±3.0%/±0.5%

80mV pk-pk/300mV pk-pk/120mV pk-pk (note 1)

MPT120 Triple Series

Model Number

Voltage Tolerance

Over Voltage Protection

Over Current Protection

Rise Time

MPT120-A ±2.0%/+8,-6%/±5.0% 115~135% Output shutdown when activated auto recover 110~150% rated output power, auto recover 500mS, 20mS/230VAC / 1200mS. 30mS/115VAC
MPT120-B ±2.0%/±6.0%/±5.0% 115~135% Output shutdown when activated auto recover 110~150% rated output power, auto recover 500mS, 20mS/230VAC / 1200mS. 30mS/115VAC
MPT120-C ±2.0%/+6, -7%/±5.0% 115~135% Output shutdown when activated auto recover 110~150% rated output power, auto recover 500mS, 20mS/230VAC / 1200mS. 30mS/115VAC
MPT120-D ±2.0%/+8, -6%/±5.0% 115~135% Output shutdown when activated auto recover 110~150% rated output power, auto recover 500mS, 20mS/230VAC / 1200mS. 30mS/115VAC

Input Specifications

Input Voltage Range

90-264VAC / 127-370vdc

Frequency Range

47-440 Hz

Inrush Current

<22 Amps @ 115VAC Input, 45 Amps @ 230VAC (cold start)

Input Current

2.9 Amps max @ 115VAC 1.7 Amps max @ 230VAC

Leakage Current

0.18mA / 230 VAC

Harmonic Input Current

Compliance to EN61000-3-2,-3

Environmental Specifications

Operating Temperature

-20°C to +70°C (See Derate Curve)

Storage Temperature

-40°C to +85°C, 10~90% RH

Relative Humidity

20 to +90% RH

EMI

Compliance to EN55011B,EN55022B

MTBF

262.1KHrs min, MIL-HDBK-217F(25°C)

Vibration

10~500Hz, 2G10min./1cycle, period for 60min. each along X, Y, Z axes

EMS

Compliance to EN61000-4-2,3,4,5,6,8,11; ENV50204, EN55024, EN60601-1-2, medical level, criteria A

General Specifications

Isolation

4000VAC

Isolation Input-Ground

1500VAC

Isolation Output-Ground

500VAC

Insulation Resistance

≥100MΩ

Efficiency

75% typ.

Safety Approvals

UL60601-1 / UL60950-1 TUV EN60601-1 / TUV EN60950-1 IEC60601-1 approved

Output Specifications

Output Voltage

3.3vdc

Output Amps

24 Amps

Minimum Load Requirement

0%

Load Regulation

±3.0%

Line Regulation

±1.0%

Temperature Coefficient

0.04% / °C (0-50°C)

Ripple and Noise

80mV pk-pk (note 1)

Voltage Tolerance

±3.0%

Peak Load

108% of output min. (note 3)

Rise Time

800mS, 20mS/230VAC / 2000mS, 50mS/115VAC

Hold Up Time

80mS @ 230VAC 14mS @ 115VAC

Physical Specifications

Size (mm)

177.8 x 107.95 x 35.5 mm

Size (Inches)

7” x 4.25” x 1.4”

Weight (g)

550g

Weight (oz)

14.40oz

Notes

1. Ripple and Noise are measured at 20MHz of bandwidth by using a 12” twisted pair-wire terminated with a 0.1uf & 47 uf parallel capacitor.

2. Tolerance: includes set up tolerance, line regulation and load regulation.

3. 33% Duty cycle max within every 30S. Average output power should not exceed the rated power.

All specifications are typical at nominal input, full load, and 25°C unless otherwise noted

Astrodyne products are not authorized or warranteed for use as critical components in life support systems, equipment used in hazardous environments, nuclear controls systems, or other mission-critical applications.