Order by phone: 1-800-823-8082 / +1-508-964-6300 M-F 8:30am to 5:30pm EST

MPS200 200w Power Supplies

Universal AC input

Low leakage current ≤ 0.18mA

Protections: Short circuit / Overload / Over voltage / Over temperature

Fixed switching frequency at 100KHz

Free air convection for 140W and forced air convection for 200W

UL60601-1 medical safety approved

Model Number

Output Voltage

Output Amps

Minimum Load Requirement

Load Regulation

Line Regulation

Ripple and Noise

Voltage Tolerance

MPS200-3.3

3.3 Volts(DC)

40 Amps

0%

±1.0%

±0.5%

80mV pk-pk (note 1)

±2.0%

MPS200-5

5 Volts(DC)

40 Amps

0%

±1.0%

±0.5%

80mV pk-pk (note 1)

±2.0%

MPS200-12

12 Volts(DC)

16.7 Amps

0%

±1.0%

±0.5%

100mV pk-pk (note 1)

±2.0%

MPS200-15

15 Volts(DC)

13.4 Amps

0%

±1.0%

±0.5%

100mV pk-pk (note 1)

±2.0%

MPS200-24

24 Volts(DC)

8.4 Amps

0%

±1.0%

±0.5%

150mV pk-pk (note 1)

±1.0%

MPS200-48

48 Volts(DC)

4.2 Amps

0%

±1.0%

±0.5%

200mV pk-pk (note 1)

±1.0%

MPS200 Single Series

Input Specifications

Input Voltage Range

90-264VAC / 127-370 Volts(DC)

Frequency Range

47-440 Hz

Inrush Current

<25 Amps @ 115VAC, Input 40 Amps @ 230VAC (cold start)

Input Current

3.5 Amps max @ 115VAC 1.6 Amps max @ 230VAC

Leakage Current

heart leakage < 0.18mA / 264VAC; patient leakage < 0.1mA / 264VAC

Power Factor Correction

PF>0.95/230VAC PF>0.98/115VAC

Harmonic Input Current

Compliance to EN61000-3-2,-3

Environmental Specifications

Operating Temperature

0.04% / °C (0-50°C)

EMC

Storage Temperature

-40°C to +85°C, 10~90% RH

Relative Humidity

20 to +90% RH

EMI

Compliance to EN55011, EN55022 Class B

MTBF

547KHrs min, MIL-HDBK-217F(25°C)

Vibration

10~500Hz, 2G10min./1cycle, period for 60min. each along X, Y, Z axes

EMS

Compliance to EN61000-4-2,3,4,5,6,8,11; ENV50204

General Specifications

Isolation

4000VAC

Isolation Input-Ground

1500VAC

Isolation Output-Ground

1500VAC

Insulation Resistance

≥100MΩ

Efficiency

81% typ.

Safety Approvals

UL60601-1 / UL60950-1 TUV EN60601-1 / TUV EN60950-1 IEC60601-1 approved

Output Specifications

Output Voltage

3.3 Volts(DC)

Output Amps

40 Amps

Minimum Load Requirement

0%

Load Regulation

±1.0%

Line Regulation

±0.5%

Temperature Coefficient

0.04% / °C (0-50°C)

Ripple and Noise

80mV pk-pk (note 1)

Voltage Tolerance

±2.0%

Over Voltage Protection

5Volts only:115~135%, shutdown

Over Current Protection

Hiccup type 120~160% auto recover

Over Temperature Protection

TSW1 Temp ? 95°C, shutdown auto recover

Peak Load

120% of output min. (note 3)

Rise Time

1000mS, 20mS / 230VAC 3000mS, 20ms / 115VAC

Hold Up Time

16mS @ 230VAC 16mS @ 115VAC

Remote shutdown

RC+/RC-: 0~0.8Vots(DC) = Power On 4~10Volts(DC) = Power Off

DC Voltage Adjust (typ.)

-5%~10% rated output volt by potential meter

Physical Specifications

Size (mm)

177.8 x 107.2 x 35.5 mm

Size (Inches)

7” x 4.22” x 1.4”

Weight (g)

660g

Weight (oz)

23.28oz

Notes

1. Ripple and Noise are measured at 20MHz of bandwidth by using a 12” twisted pair-wire terminated with a 0.1uf & 47 uf parallel capacitor.

2. Tolerance: includes set up tolerance, line regulation and load regulation.

All specifications are typical at nominal input, full load, and 25°C unless otherwise noted

Astrodyne products are not authorized or warranteed for use as critical components in life support systems, equipment used in hazardous environments, nuclear controls systems, or other mission-critical applications.