Order by phone: 1-800-823-8082 / +1-508-964-6300 M-F 8:30am to 5:30pm EST

MMK150 150w Switch Mode Medical Power Supplies

Medical Grade 150 Watt Enclosed Switching Power Supplies

Universal 88-264VAC Auto Range Input

Compact Footprint

High Efficiency

Regulated Outputs

4000V Isolation

Built in EMI Filter

Model Number

Output Voltage

Output Amps

MMK150S-5

5 VDC

30A

MMK150S-12

12VDC

12.5A

MMK150S-15

15VDC

10A

MMK150S-24

24VDC

6A

MMK150S series

Input Specifications

Input Voltage Range

88-132/176-264 VAC Auto Voltage Select

Start Up Time

? 1.0 sec, typ

Frequency Range

47-63 Hz

Inrush Current

20A @ 100VAC *, 40A @ 200VAC *

General Specifications

Isolation

5656VDC 1 Sec. max.

Isolation Input-Ground

2121VDC 1 Sec. max.

Isolation Output-Ground

707VDC 1 Sec. max.

Insulation Resistance

78%, typ

Efficiency

100Khz, (fixed, typical)

Switching Frequency

UL60601-1, EN60601-1, IEC60601-1

Environmental Specifications

Operating Temperature

0 to +60°C(see derate)

Storage Temperature

-25 to +85°C *

Relative Humidity

10% to 90%, non-cond *

EMI

EN55011, Class B, (Designed to meet)

MTBF

100,000 Hrs Mil Std 217, 25°C

Output Specifications

Output Voltage

5 VDC

Output Amps

30A

Load Regulation

± 0.5%, typ (NL-FL)

Line Regulation

± 0.5%, typ (NL-FL)

Temperature Coefficient

± 0.03%/°C

Ripple and Noise

100mV, typ

Over Voltage Protection

130%, Latching, Re-power *

Short Circuit Protection

160%, Indefinite

DC Voltage Adjust (typ.)

± 6%, typ

Hold Up Time

20mS, typ (Nom I/P, FL)

Physical Specifications

Notes

Size (Inches)

6” x 3.5” x 1.75”

Construction

Enclosed

Weight (g)

630g

Weight (oz)

21.9oz

* These are stress ratings. Exposure of the devices to any of these conditions may adversely affect long term reliability. Proper operation under conditions other than the standard operating conditions is neither warranteed nor implied.

All specifications are typical at nominal input, full load, and 25°C unless otherwise noted

Astrodyne products are not authorized or warranteed for use as critical components in life support systems, equipment used in hazardous environments, nuclear controls systems, or other mission-critical applications.